제품브랜드별

FLUKE

FLUKE-724

INFO
제조사 FLUKE
원산지 US
브랜드 FLUKE
모델 FLUKE-724
판매가격 0원

선택된 옵션

  • FLUKE-724 (+0원)
ICON 견적요청 ICON 카달로그

FLUKE 724 TC/RTD 온도 캘리브레이터는 다양한 플랜트 현장에 시설되어 있는 온도 트랜스미터와 인디게이터의 정확한 동작을 위한 점검과 교정에 필요한 신호를 측정하고 공급하는 어플리케이션에 적용되는 온도 측정, 발생 장비 입니다.

 

-소스/측정 TCs, RTDs, 전압, 저항

-루프파워 공급시 mA측정

-25%, 100% 스텝,자동 스텝, AUTO ramp

 

Measurement Accuracy

Voltage DC

30.000 V

 0.02%+ 2 counts (upper display)

20.000 V

 0.02%+ 2 counts (lower display)

100.00 mV

 0.02%+ 2 counts

-10.00 mV to 75.00 mV

 0.025 % + 1 count (via TC connector)

Current DC

24.000 mA

 0.02%+ 2 counts

Resistance

0.0 to 400.0 Ω

 0.1Ω (4-wire), 0.15Ω (2- and 3-wire)

401 to 1500 Ω

 0.5Ω (4-wire), 1Ω (2- and 3-wire)

1500 to 3200 Ω

 1Ω (4-wire), 1.5Ω (2- and 3-wire)

 

 

Source Accuracy

Voltage DC

100.00 mV

 0.02%+2 counts

10.000 V

 0.02%+2 counts

-10.00 mV to 75.00 mV

 0.025 % + 1 count (via TC connector)

Resistance

15.0 to 400.0 Ω

 0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA)

401 to 1500 Ω

 0.5 Ω(excitation current 0.05 to 0.8 mA)

1500 to 3200 Ω

 1 Ω(excitation current 0.05 to 0.4 mA)

 

 

RTDs and Thermocouples

Measure Accuracy:

NI-120

 0.2 °C

PT-100 (385)

 0.33 °C

PT-100 (393)

 0.3 °C

PT-100 (JIS)

 0.3 °C

PT-200 (385)

 0.2 °C

PT-500 (385)

 0.3 °C

PT-1000 (385)

 0.2 °C

Resolution:

 0.1 °C

J:

 0.7 °C

K:

 0.8 °C

T:

 0.8 °C

E:

 0.7 °C

R:

 1.8 °C

S:

 1.5 °C

B:

 1.4 °C

L:

 0.7 °C

U:

 0.75 °C

N:

 0.9°C

Resolution

 J, K, T, E, L, N, U: 0.1 °C, 0.1 °F B, R, S: 1 °C, 1 °F

XK

 0.6°C

BP

 1.2°C

Source Accuracy:

NI-120

 0.2 °C

PT-100 (385)

 0.33 °C

PT-100 (393)

 0.3 °C

PT-100 (JIS)

 0.3 °C

PT-200 (385)

 0.2 °C

PT-500 (385)

 0.3 °C

PT-1000 (385)

 0.2 °C

Resolution:

 0.1 °C

Note:

 Accuracy stated for 4-wire measurement.

J:

 0.7 °C

K:

 0.8 °C

T:

 0.8 °C

E:

 0.7 °C

R:

 1.4 °C

S:

 1.5 °C

B:

 1.4 °C

L:

 0.7 °C

U:

 0.75 °C

N:

 0.9 °C

Resolution

 J, K, T, E, L, N, U: 0.1 °C, B, R, S: 1 °C

XK

 0.6°C

BP

 1.2°C

 

 

Specifications

Ramp Functions

Source functions:

 Voltage, current, resistance, frequency, temperature

Ramps

 Slow ramp, Fast ramp, 25% step-ramp

Loop Power Function

Voltage:

 24V

Accuracy:

 10%

Maximum current:

 22 mA, short circuit protected

Step Functions

Source functions:

 Voltage, current, resistance, frequency, temperature

Steps

 25% of range, 100% of range

 

 

Environmental Specifications

Operating Temperature

 -10ºC to 55ºC

Storage Temperature

 -20ºC to 71ºC

Humidity (Without Condensation)

90%

 10ºC to 30ºC

75%

 30ºC to 40ºC

45%

 40ºC to 50ºC

35%

 50ºC to 55ºC

 

 

Safety Specifications

Safety Rating

 CSA C22.2 No. 1010.1:1992

EMC

 EN50082-1:1992 and EN55022:1994 Class B

 

 

Mechanical & General Specifications

Size

 96 x 200 x 47 mm

Weight

 650 g

Batteries

 Four AA alkaline batteries.

Warranty

 Three years

Battery Life

 25 hours typical

Shock & Vibration

 Random, 2G, 5-500 Hz

 

Model Name

Description

Fluke 724

Temperature Calibrator including:

AC70A Test Clips

One pair of stackable test leads

TL75 Test Leads

CE and CSA Markings

Manuals in English, French, German, Spanish, Italian, Portugese, Korean, Chinese and Japanese

Users manuals on CD-ROM

NIST- traceable calibration report and data